Abstract

The aim of this work is to provide the characteristics of MPPC arrays with an active area of 6x6 mm2 in gamma-ray spectrometry with 5x5x5mm3 CsI:TI scintillator at a wide temperature range. The measurements presented in this work covered: characteristics of breakdown voltage, selection of MPPC optimum operating voltage, verification of excess noise factor, measurement of energy resolution and the number of photoelectrons per 1 MeV. All measurements were carried out as a function of temperature and compared with similar results obtained previously with CsI:TI crystal coupled to a Large Area Photodiode (LAAPD).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.