Abstract
Bioresorbable materials for implants have become increasingly researched over the last years. The bone–implant-interfaces of three different implant materials, namely a new bioresorbable magnesium alloy, a new self-reinforced polymer implant and a conventional titanium alloy, were tested using various methods: push-out tests, SEM and EDX analyses as well as surface analyses based on stereoscopic 3D pictures were conducted. The fracture energy is proposed as a very significant reference value for characterizing the mechanical performance of a bone–implant system. By using a video-extensometer system instead of, as is commonly done, tracking the movement of the crosshead in the push-out tests, the accuracy of measurement could be increased.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of the Mechanical Behavior of Biomedical Materials
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.