Abstract
Bonner sphere (BS) sets which use activation foils as the central thermal neutron sensor have advantages over active BS systems in certain environments, for example, pulsed fields, or fields with high photon components. In such environments, they may be the only type of neutron spectrometer which can be used. This paper describes work, using both measurements and calculations, to validate the response functions for a BS set based on gold activation foils. As an illustration of the use of such a system, a measurement is described of the contaminant neutron spectrum in the treatment room of a 21 MV hospital linear accelerator providing photon beams for radiotherapy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.