Abstract
A simple, low cost TLC scanner system for the characterization of γ-emitting radiopharmaceuticals is described. The TLC γ scanner has a linear response range of ∼0.074 Mbq up to ≥2.94 Mbq. The accuracy and reproducibility of measurement is better than ±5% (relative). The utility of this TLC scanner system is demonstrated by application to the radioanalytical evaluation of99mTc(dmpe)2Cl 2 + preparations. A comparison with HPLC radioanalytical separations is made.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.