Abstract

The characteristics of n-type diodes (linearity, temperature, dose rate, radiation damage response, directional dependence, output factors, wedge factors and percentage depth dose determinations) were investigated. Subsequently, the diodes were used clinically for in vivo dose verification during external megavoltage photon beam therapy. It has been shown that n-type diodes are easy to use and the results obtained are comparable to those reported for p-type diodes. On most occasions, n-type diodes can be used without any additional correction factors apart from regular monthly calibration. There is good agreement between the uncertainty limits estimated from the diodes' characteristics and those obtained on the basis of 2261 patient measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.