Abstract

We are developing a human brain PET scanner prototype compatible with MRI based on monolithic scintillator crystals, APD matrices and a dedicated ASIC front-end readout. In this work we report on the performance of individual detector modules and on the operation of such modules in PET coincidence. Results will be presented on the individual characterization of detector blocks and its ASIC front-end readout, with measured energy resolutions of 13% full-width half-maximum (FWHM) at 511 keV and spatial resolutions of the order of 2 mm FWHM. First results on PET coincidence performance indicate spatial resolutions as good as 2.1 mm FWHM for SSRB/FBP reconstruction of tomographic data obtained using a simple PET demonstrator based on a pair of monolithic detector blocks with ASIC readout.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.