Abstract

We have developed prototype arrays of metallic magnetic calorimeters for applications in X-ray astronomy. Each pixel consists of an all-gold X-ray absorber in good thermal contact to a gold-erbium paramagnetic thin film thermometer that is operated in the temperature range of 30–100 mK. The para-magnetic response is coupled to a SQUID amplifier. We have characterized pixels in an array and observed the expected temperature dependence of the magnetization and heat capacity. We have demonstrated a full width at half maximum energy resolution of 1.7 \(\pm \) 0.1 eV at 6 keV and have also read out these devices using time-division multiplexing.