Abstract
Fifteen electrolytic flow-through cells belonging to three different design groups (non-membrane, thin-layer and tubular) have been constructed and characterized in this work. Atomic absorption spectrometry with a quartz tube atomizer has been employed as a detector and Se chosen as a model analyte. At first optimization of relevant working parameters was performed and basic analytical characteristics were evaluated under the optimum experimental conditions for all cells. Mutual comparisons of the electrolytic cells were performed thereafter and 5 cells with best performance were selected for further evaluation. Finally, a 75Se radioactive indicator was employed to quantify the generation efficiency of selane as well as to visualize the spatial distribution of the analyte in the apparatus and track potential analyte losses. With respect to all relevant requirements the best performance was found for the thin-layer electrolytic cell (225 mm3 cathode chamber volume) with efficiency of electrochemical hydride generation reaching 75% and a limit of detection (LOD) of 0.31 ng mL−1 Se. A generation efficiency of 78% and an LOD of 0.40 ng mL−1 Se were obtained for the tubular cell (43 mm3 cathode volume). High generation efficiency, fast response and low cathode chamber volume make the tubular cell an ideal candidate even for flow-injection and speciation analysis measurements.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.