Abstract
Plasmatic chlorophacinone is commonly measured with liquid chromatographic assay, which convenient but lacks sensitivity and selectivity and usually requires ion pair reagents to reduce the chromatographic tailed peak. In this paper, a novel method using eluent generator reagent-free ion chromatography coupled with electrospray ionization ion trap mass spectrometric detection for the determination of chlorophacinone in plasma has been developed. After samples were extracted with 10% (v/v) methanol in acetonitrile and cleaned by solid-phase extraction, chromatographic separation was performed on an IonPac AS11 analytical column (250 x 4.0 mm) using 40.0 mmol/L KOH containing 10% (v/v) methanol as organic modifier. Quantification was performed by negative electrospray ionization in multiple reaction monitoring mode. The transition m/z 373 --> 201 was for the quantification ion; the transitions m/z 373 --> 172 and m/z 373 --> 145, as well as the isotope ions m/z 375 and m/z 203, were for the qualitative ions. All the method parameters were validated. It was confirmed that this method can be used in clinical diagnosis and forensic toxicology.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.