Abstract
The purpose of this study was to demonstrate and correct the cupping effect artefact that occurs owing to the presence of beam hardening and scatter radiation during image acquisition in cone beam CT (CBCT). A uniform aluminium cylinder (6061) was used to demonstrate the cupping effect artefact on the Planmeca Promax 3D CBCT unit (Planmeca OY, Helsinki, Finland). The cupping effect was studied using a line profile plot of the grey level values using ImageJ software (National Institutes of Health, Bethesda, MD). A hardware-based correction method using copper pre-filtration was used to address this artefact caused by beam hardening and a software-based subtraction algorithm was used to address scatter contamination. The hardware-based correction used to address the effects of beam hardening suppressed the cupping effect artefact but did not eliminate it. The software-based correction used to address the effects of scatter resulted in elimination of the cupping effect artefact. Compensating for the presence of beam hardening and scatter radiation improves grey level uniformity in CBCT.
Published Version (Free)
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.