Abstract
Abstract The high temperature thermoluminescence (HTTL) in the glow curve of LiF:Mg,Ti (TLD-100) is of continuing interest due to the dependence of the HTTL on ionization density. Following high ionization density irradiation, the intensity of the HTTL relative to peak 5 is much greater than the same ratio following low ionization density irradiation. Special attention has focused on the details of the linearity/supralinearity of the HTTL dose response at levels of dose between 10 and 500 mGy and whether the batch history (previous irradiation and annealing) affects the HTTL dose response? In this paper we demonstrate that batch identity leads to differences of ∼100% in the degree of the supralinearity of the dose response of peak 7 between two groups of TLD-100 chips. The results suggest that there may be significant non-universalities in the HTTL dose response and that further research is required to provide a sound basis for the use of the HTTL of TLD-100 in dosimetric applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.