Abstract

Asymptomatic paroxysmal atrial fibrillation (AF) is often found in patients implanted with cardiac implantable electronic devices (CIEDs). Second-generation atrial antitachycardia pacing (A-ATP) is effective in managing AF in patients implanted with CIEDs. The purpose of this study was to evaluate the efficacy and safety of A-ATP in patients implanted with CIEDs. This was a single-center retrospective study involving 91 patients (male 46 patients, mean age 74 ± 9years) implanted with Reactive A-ATP equipped devices (84 patients with pacemakers, 6 with ICDs, and 1 with a CRT-D). The AF burden, rate of AF termination, and details of the activation of the A-ATP were analyzed in each patient. During a mean follow-up period of 21 ± 13months, A-ATP was activated in 45 of 91 patients (49.5%). No patients had adverse events. Although the efficacy of the A-ATP varied among the patients, the median rate of AF termination was 44%. In comparison to the A-ATP start time, "0min" had a higher AF termination rate by the A-ATP (39.4% vs. 24.4%, P = 0.011). The rate of termination by the A-ATP was high for AF with a long cycle length and a relatively regular rhythm. A-ATP successfully terminated AF episodes in some patients implanted with CIEDs. The optimal settings of the A-ATP will be determined in future studies.