Abstract

Alpha spectrometry is an extremely useful and sensitive for detection of alpha-emitting nuclides. Contamination of the silicon detectors for low-level alpha spectrometry by recoil nuclides is a serious problem in the measurement of alpha emitters decaying to daughter nuclides with short half-lives. This unwanted contamination leads to decreased measurement sensitivity causing a degradation of the limit of detection. The simplest method to prevent this radioactive contamination of detector is to use a catcher film between the alpha source and the detector. In this work we describe the obtaining of the thin formvar films as stopper foils for recoil nuclei and we investigated the influence of these films on alpha spectrometry parameters, as energy shift (~30 keV) and resolution (~7%). No significant deterioration of the alpha spectrometry parameters was observed when using thin formvar films. Using the ASTAR web databases, which calculate stopping powers for alpha particles, the thickness of formvar films was estimated to be about 5.355 × 10−5 g/cm2. The measurements were performed with an ORTEC SOLOIST alpha spectrometer with PIPS detector.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.