Abstract

The purpose of this study is to characterize the dosimetric properties and accuracy of a novel treatment platform (Edge radiosurgery system) for localizing and treating patients with frameless, image‐guided stereotactic radiosurgery (SRS) and stereotactic body radiotherapy (SBRT). Initial measurements of various components of the system, such as a comprehensive assessment of the dosimetric properties of the flattening filter‐free (FFF) beams for both high definition (HD120) MLC and conical cone‐based treatment, positioning accuracy and beam attenuation of a six degree of freedom (6DoF) couch, treatment head leakage test, and integrated end‐to‐end accuracy tests, have been performed. The end‐to‐end test of the system was performed by CT imaging a phantom and registering hidden targets on the treatment couch to determine the localization accuracy of the optical surface monitoring system (OSMS), cone‐beam CT (CBCT), and MV imaging systems, as well as the radiation isocenter targeting accuracy. The deviations between the percent depth‐dose curves acquired on the new linac‐based system (Edge), and the previously published machine with FFF beams (TrueBeam) beyond Dmax were within 1.0% for both energies. The maximum deviation of output factors between the Edge and TrueBeam was 1.6%. The optimized dosimetric leaf gap values, which were fitted using Eclipse dose calculations and measurements based on representative spine radiosurgery plans, were 0.700 mm and 1.000 mm, respectively. For the conical cones, 6X FFF has sharper penumbra ranging from 1.2−1.8 mm (80%‐20%) and 1.9−3.8 mm (90%‐10%) relative to 10X FFF, which has 1.2−2.2 mm and 2.3−5.1 mm, respectively. The relative attenuation measurements of the couch for PA, PA (rails‐in), oblique, oblique (rails‐out), oblique (rails‐in) were: −2.0%, −2.5%, −15.6%, −2.5%, −5.0% for 6X FFF and −1.4%, −1.5%, −12.2%, −2.5%, −5.0% for 10X FFF, respectively, with a slight decrease in attenuation versus field size. The systematic deviation between the OSMS and CBCT was −0.4±0.2 mm, 0.1±0.3 mm, and 0.0±0.1 mm in the vertical, longitudinal, and lateral directions. The mean values and standard deviations of the average deviation and maximum deviation of the daily Winston‐Lutz tests over three months are 0.20±0.03 mm and 0.66±0.18 mm, respectively. Initial testing of this novel system demonstrates the technology to be highly accurate and suitable for frameless, linac‐based SRS and SBRT treatment.PACS number: 87.56.J‐

Highlights

  • Since the term “stereotactic radiosurgery” was coined by Lars Leksell in 1951, there have been many technological, biological, and clinical advances in the field of stereotactic radiosurgery.[1,2,3,4] The accuracy of linear accelerators has been improved significantly since the 1980s(5-7) and linac-based radiosurgery has been widely adopted over the subsequent decades

  • AAPM task group report No 45 “AAPM Code of Practice for Radiotherapy Accelerators” recommendations were followed for commissioning tasks.[22]. Selection of different detectors for water phantom measurements were based on AAPM task group report No 106 and small field dosimetry specification[23] (Table 1)

  • Field sizes ranged from 1 × 1 cm2 to 40 × 40 cm2 which were determined by the jaw

Read more

Summary

Introduction

Since the term “stereotactic radiosurgery” was coined by Lars Leksell in 1951, there have been many technological, biological, and clinical advances in the field of stereotactic radiosurgery.[1,2,3,4] The accuracy of linear accelerators (linacs) has been improved significantly since the 1980s(5-7) and linac-based radiosurgery has been widely adopted over the subsequent decades. The new couch (PerfectPitch) supports six degrees of freedom (6DoF) corrections from multiple imaging modalities for precise patient setup. The flat panel imager is designed with a greater dynamic range, faster image readout rate, and a larger active area. This technology has a stereotactic accessory package which includes conical cones ranging in diameter from 4 to 17.5 mm. We describe a comprehensive commissioning process suitable for modern, linac-based SRS/ SBRT with focus on the characterization of beam parameters, conical cones, 6DoF couch, dosimetric verification, and integrated end-to-end tests of this new technology

Objectives
Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call