Abstract

With this rise in surgical procedures it is important to focus on the mobility and safety of the patient and reduce the infections that are associated with hip replacements. We examine the mechanical properties of gentamicin sulphate as a model antimicrobial layer for titanium-alloy based prosthetic hips to help prevent methicillin-resistant Staphylococcus aureus infection after surgery. A top layer of poly(lactic-co-glycolic acid) is added to maintain the properties of the gentamicin sulphate as well as providing a drug delivery system. Through the use of nanoindentation and micro-scratch techniques it is possible to determine the mechanical and adhesive properties of this system. Nanoindentation determined the modulus values for the poly(lactic-co-glycolic acid) and gentamicin sulphate materials to be 8.9 and 5.2GPa, respectively. Micro-scratch established that the gentamicin sulphate layer is strongly adhered to the Ti alloy and forces of 30N show no cohesive or adhesive failure. It was determined that the poly(lactic-co-glycolic acid) is ductile in nature and delaminates from the gentamicin sulphate layer of at 0.5N.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.