Abstract
A number of state distribution temperatures were measured in an effort to characterize a 100 W 13.56 MHz rf He plasma established in a FAPES (Furnace Atomization Plasma Emission Spectrometry) source. Excitation temperatures derived from a Boltzmann plot of He I and Fe I lines were in reasonable agreement, i.e. 3300 ± 180 and 2920 ± 180 K, respectively. An excitation temperature measured with Fe II lines was substantially larger than the above values, being 7610 ± 450 K. A kinetic or Doppler temperature based on measurement of the width of the Be I 234.861 nm line profile was 7030 ± 1150 K whereas the He ionization temperature derived from the measured electron number density was 10920 ± 1900 K. It is clear that complete LTE does not exist in this source
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.