Abstract

In this study conducted to evaluate the collimator 4 mm, 8 mm, 16 mm to determine the level of precision Gamma Knife Perfexion there three years and eight months has not carried out measurements after the first year and verify the alignment of sector 1-8. Measurement with three axes (x-axis, y-axis, z-axis) using the film Gafcromic EBT-3, which will be signaled to a central point in the film to be measured and given a dose of 5 Gy for 10 minutes and then scanned and analyzed using ImageJ softwere 1.50 and dose rate in film comparison with measurent dose rate using electrometer. Alignment of Patient-Positioning System (PPS) with Radiation Focal Point (RFP) determination of quality control testing each collimator 4 mm in the standard test with < 0.1 and > 0.4 mm passive voice AAPM 54. The results of the standard are still in conformity item 0.21 mm for 4 mm and corelasi dose rate film and measurent using electrometer collimator 4 mm = 0.965, 8 mm = 0.964, 16 mm = 0.959.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.