TruView™ is a new water-equivalent reusable Fricke gel dosimeter based on MethylThymolBlue reactive dye. Details of the characterisation of the TruView™ MTB gel dosimeter by spectrophotometric measurements and of its reading with the Optical-CT Scanner Vista™ are described. In this study, the different parameters influencing TruView™ dose response have been studied and its performances have been compared to chamber and diodes measurements. This gel presents a linear response with dose up to 20 Gy, independent in the investigated range of photon beam energy and dose rate and also a good intra-batch uniformity. Ions diffusion into the matrix homogenizes the gel after a week, losing dosimetric information but allowing a new irradiation to be performed. However, auto-oxidation happens before and after irradiation, degrading the dosimeter response and stability. Storage and reading conditions affect the response as well.