Abstract
The objective of this study was to characterise thermoluminescent (TLDs) and optically stimulated luminescent dosimeters (OSLDs) at low X-ray energies and estimate the eye lens (DL), thyroid (DT) and mean glandular (DG) doses received during Full-Field Digital Mammography (FFDM) and Digital Breast Tomosynthesis (DBT). The dosimeters were characterised in mammography energies. DL, DT and DG were estimated in FFDM and DBT mode taping dosimeters on the skin of the thyroid gland and on the left eye lens of an Alderson phantom. Dosimeters were also placed on the top of a NORMI PAS phantom simulating a compressed breast. The accuracy, precision and lower limit of detection (LLD) for TLDs and OSLDs were 5 and 8%, 6 and 3%, and 38 and 11 μSv, respectively. The linearity of the kerma response had an R2 > 0.99 and energy dependence was lower than 40%. DT ranged from 0.40 to 2.87 μGy for FFDM and 1.27 to 5.99 μGy for DBT. DG was between 0.50 and 1.27 mGy for FFDM and 1.07 and 1.60 mGy for DBT. DL was below the LLD. Dosimeters showed good performance. DG values were lower than those found in the literature, whereas DT value agreed with references. Differences between DG and DT determined with OSLDs and TLDs were lower than 10% and 200%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.