Abstract
The imaging performance of an XPS instrument employing a spherical mirror electron energy analyser has been characterised by measuring the peak position, full width at half maximum (FWHM), and lineshape, at every pixel in the image, for different modes of operation. Changes in these parameters have been identified and recommendations made for quantification of, and chemical state determination from, spectrum image data sets.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.