Abstract
Modern stereotactic radiation therapy modalities utilize small beams and large dose gradients to deliver radiation in few fractions, reducing the possibility to correct for mistakes during the treatment process. Therefore, in order to ensure best possible treatment for the patient, quality assurance for such treatments necessitates a stable, linear, and sensitive radiation detector with high spatial resolution and radiation hardness. In this work, two silicon detector arrays with high spatial resolution have been characterized by 6 MV and 18 MV medical LINAC irradiation, and 5.5 MeV He2+ heavy ion microprobe. A maximum discrepancy of 0.6 mm in field size has been found when comparing to two-dimensional radiochromic film dose profile, and charge collection efficiency obtained by means of ion beam induced charge collection (IBICC) is 66% when operating the array in photovoltaic mode. Radiation damage study by photons and photoneutrons is presented.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.