Abstract

The authors have previously published a method of beam hardening correction for X-ray microtomography that allows accurate determination of mineral concentration in hard tissue. This uses a multi-element calibration carousel for making experimental X-ray attenuation measurements to optimise the parameters in a model of X-ray transmission and detection. From this, a calibration curve is generated to convert the X-ray polychromatic X-ray projection data to that which would be expected with monochromatic X-rays. Furthermore, the calibration curve can be modified according to the known composition of the specimen. Here, this is tested using different materials to assess its accuracy in determining their density. Generally, errors in density of around 1–2% were obtained. When modelling the conventional use of an aluminium step wedge, much larger errors are seen. The technique can be used to measure and map density or concentration in cases where this would be difficult with other methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.