Abstract

A study was undertaken to evaluate the response of the ENEA Photon Personal Dosemeter, slightly modified in design, in terms of Hp(0.07, a) and Hp(10, a). A filtration of 20 and 270 mg.cm-2 respectively for the two LiF(Mg,Cu,P) detectors employed was adopted according to a preliminary analytical simulation. The new dosemeter can measure both Hp(0.07, a) with a maximum uncertainty of ±4% in the range from 10 to 200 keV and Hp(10, a) with a maximum uncertainty of ±15% in the range from 10 keV to 1.25 MeV, using the detector responses together with a correction function with respect to their ratio. Moreover, tests were performed to evaluate the dosemeter angular response (0(, 20(, 40(, and 60() obtaining good results according to international recommendations. All irradiation tests were performed at SSL of ENEA Institute for Radiation Protection in Bologna using the ISO water phantom (30 x 30 x 15 cm3) and standard ISO filtered X ray beams as well as 137Cs and 60Co sources. This work represents a preliminary type test of the ENEA photon personal dosemeter and allows its introduction in routine dosimetry practice in accordance with the ICRU reference quantities.