Abstract

A test model of a centrifugal pump with constant cross-section channels in the rotor was developed using 3D computer simulation. The channels are shaped as a logarithmic spiral. The geometry of the flow channel providing optimal flow in the nominal operating mode (flow, 5 L/min; pressure drop, 100 mm Hg) was calculated. In addition, pump operation conditions in the extracorporeal membrane oxygenation (ECMO) mode were considered. The main requirements imposed on the developed test model were those of meeting the allowable shear stress threshold and minimizing the stagnation and flow recirculation zones.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.