Abstract
A local absorbed dose calorimeter with certain novel features has been designed and constructed of A-150 tissue-equivalent (TE) plastic. The radiation absorption properties of this material and the relative absence of core impurities make the calorimeter suitable for use in a wide variety of radiation beams. The low thermal diffusivity of A-150 plastic led to the development of a unique spiral electrical calibration heater that has a mass of only 0.1% that of the core. The calorimeter can be calibrated in the quasiadiabatic or the heat-loss-compensated mode to test for possible effects caused by temperature gradients. The details of construction and operation are described.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.