Abstract

A local absorbed dose calorimeter with certain novel features has been designed and constructed of A-150 tissue-equivalent (TE) plastic. The radiation absorption properties of this material and the relative absence of core impurities make the calorimeter suitable for use in a wide variety of radiation beams. The low thermal diffusivity of A-150 plastic led to the development of a unique spiral electrical calibration heater that has a mass of only 0.1% that of the core. The calorimeter can be calibrated in the quasiadiabatic or the heat-loss-compensated mode to test for possible effects caused by temperature gradients. The details of construction and operation are described.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.