Abstract

Concerns about the potential adverse health effects of perchlorate at concentrations below the minimum reporting level (MRL) of US Environmental Protection Agency (EPA) Method 314.0 (generally recognized as 4.0 μg/l) have led to an interest in increasing the sensitivity of the method. This work describes the use of 2 mm columns with a large-loop direct injection method, a column concentration technique and this concentration technique with a background reduction step, to increase the sensitivity for the analysis of trace levels of perchlorate in high ionic strength matrices. The concentrator columns studied were the Dionex TAC LP-1 and a new Dionex high capacity Cryptand concentrator column. The use of a surrogate to monitor trapping efficiency for the concentration technique and the use of confirmational columns to minimize the potential for false positives are also discussed. The large-loop direct injection method and the column concentration methods provided acceptable data when the samples were pre-treated with solid phase pretreatment cartridges. The background reduction technique did not provide acceptable data with either of the concentrator columns evaluated.