Abstract
Concerns about the potential adverse health effects of perchlorate at concentrations below the minimum reporting level (MRL) of US Environmental Protection Agency (EPA) Method 314.0 (generally recognized as 4.0 μg/l) have led to an interest in increasing the sensitivity of the method. This work describes the use of 2 mm columns with a large-loop direct injection method, a column concentration technique and this concentration technique with a background reduction step, to increase the sensitivity for the analysis of trace levels of perchlorate in high ionic strength matrices. The concentrator columns studied were the Dionex TAC LP-1 and a new Dionex high capacity Cryptand concentrator column. The use of a surrogate to monitor trapping efficiency for the concentration technique and the use of confirmational columns to minimize the potential for false positives are also discussed. The large-loop direct injection method and the column concentration methods provided acceptable data when the samples were pre-treated with solid phase pretreatment cartridges. The background reduction technique did not provide acceptable data with either of the concentrator columns evaluated.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.