Abstract

AbstractSteam sterilization processes rely on high heat transfer rates from the steam to the medical devices. To guarantee this, all non‐condensable gases (NCGs) must be removed from the steam sterilizer before sterilization. The ratios and distribution of the NCGs in the pre‐sterilization phases of both a vacuum and a non‐vacuum steam sterilizer are investigated. Two CFD models for simulation of the NCGs in the sterilizers were developed and validated by the measurements of a pressure sensor and eight thermocouples inside the steam sterilizer. Furthermore, a novel method was developed to calculate the real ratio of NCGs based on the standard measurement (EN 13060). The results show that the NCGs ratio defined in the EN 13060 did not accurately represent the real ratios of the NCGs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.