Abstract

A chemiluminescence (CL) method using flow injection (FI) was developed for the determination of uric acid based on the enhancement chemiluminescence intensity of luminol-hexacyanoferrate(III)-hexacyanoferrate(II) in the presence of cetyltrimethylammonium bromide and the uric acid species. The linear range was 7.0 x 10(-10) - 9.0 x 10(-7) M with a detection limit (3sigma) of 2.58 x 10(-10) M, which was about two orders of magnitude lower than those reported. The proposed method was used for the determination of uric acid in real samples.