Abstract

Physical parameters and performance of a new brain-dedicated tomograph (CERASPECT) were measured and compared to those of a state-of-the-art single-headed rectangular rotating gamma camera. Spatial resolution, sensitivity, scatter fraction, contrast, uniformity and image quality of the two systems were evaluated using the same radioactive source geometry. The CERASPECT's spatial resolution was 23% higher and its sensitivity was three times greater than those of the gamma camera. The contrast measured with a 15 mm cold lesion was 60% for the CERASPECT and 45% for the gamma camera. Tomographic nonuniformities were measured as the percentage integral and differential uniformity of a cylindrical uniform phantom. At low-statistics counts per slice the integral uniformity was comparable and at high-statistics counts per slice it was equal to 11% and 20% for CERASPECT and the gamma camera, respectively. Image quality is generally superior with the dedicated brain system. The use of new devices with improved physical performance allows more accurate quantitative SPECT functional brain studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.