Abstract

The present study aimed to determine the effect of femoral component materials and sterilization methods on wear properties of total knee prostheses by using a knee simulator test and retrieval analysis. The simulator test revealed that ultrahigh molecular weight polyethylene (UHMWPE) inserts had remarkably lower wear against the ceramic femoral component than against the Co-Cr femoral component. However, the retrieval study revealed no significant difference in the linear wear between the former and the latter. The alumina ceramic/UHMWPE insert combination showed a mild wear. However, whether cross-linking by gamma-ray sterilization reduces wear remained unconfirmed. In contrast, oxidative degradation and/or delamination was confirmed. Thus, we conclude that alumina ceramic/ethylene oxide gas–sterilized UHMWPE insert in a total knee prosthesis might exhibit a good wear resistance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.