Abstract
Surface mount electronic devices (like resistors, capacitors, and inductors) extracted from the portable electronic devices (mobile phones, USB drives, etc.) have been studied for dose reconstruction using luminescence techniques for radiological or nuclear emergencies. In this work, carbon coated ceramic resistors removed from the electronic instruments are analyzed using thermo luminescence (TL) technique for the retrospective dose reconstruction. TL measurements on beta irradiated ceramic resistors exhibited one major dosimetric peak at around 540 K in addition to three other low temperature peaks. A linear beta dose response has been observed from 1 Gy to 125 Gy with Minimum Detectable Dose (MDD) of 0.1 Gy. The dose response is found to be supralinear below 1 Gy. The fading studies have been carried out and investigated up to 30 days. The important kinetic parameters for the material like TL trap depth, frequency factor and order of kinetics are evaluated by deconvolution method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.