Abstract
We present a novel centrifugal microfluidic system for the rapid amplification of a Bacillus anthracis gene, with integrated pumping and thermoelectric-based heating and valving. Polymerase chain reaction is a powerful method for rapid and accurate diagnosis of infectious diseases and bio-warfare agents. The fully integrated system and disposable disc developed here automates the fluidic pumping, valving, and rapid thermocycling required for polymerase chain reaction. We demonstrate the first utilization of ice-valving in an integrated centrifugal microfluidic system as a method for sealing the thermocycling chamber to reduce fluid loss due to evaporation. The automated centrifugal system is validated via amplification of a B. anthracis gene. Fluidic thermal profiles are monitored and optimized to minimize non-specific amplification, reducing the total amplification time from 110 min down to 53 min.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.