Abstract
To evaluate the effectiveness of automated cleaning of 4 mm liposuction cannulas using ultrasonic washer machine, screening for dirt using a cannulate microscope device and the screening of viable strains of Staphylococcus aureus, with microbiological analysis. Experimental study with 14 units of 4 mm liposuction cannulas performed at the Central Sterile Services Department and at the Microbiology Laboratory of a Hospital Complex, located in Porto Alegre, RS, Brazil, from October 2018 to April 2019. Of the 14 liposuction cannulas that constituted the sample, 42.9% remained dirty despite automated cleaning, and all of them showed recovery of viable microrganisms in the microbiological laboratory. The study showed that it is not possible to ensure the cleaning efficacy of automated cleaning of 4 mm liposuction cannulas, especially due to the conformation of the material with internal spaces with accumulation of dirt.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.