Abstract
A novel double-pass instrument and its data analysis method for the measurement of central and peripheral refraction is presented and validated in a group of healthy subjects. The instrument acquires in-vivo, non-cycloplegic, double-pass, through-focus images of the eye's central and peripheral point-spread function (PSF) using an infrared laser source, a tunable lens and a CMOS camera. The through-focus images were analyzed to determine defocus and astigmatism at 0° and 30° visual field. These values were compared to those obtained with a lab-based Hartmann-Shack wavefront sensor. The two instruments provided data showing good correlation at both eccentricities, particularly in the estimation of defocus.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.