Abstract

PurposeThe results of ceramic-on-ceramic (CoC) bearing surfaces in primary total hip arthroplasty (THA) were well known. However, it was not known in revision THA. The purpose of this study is to report the results of revision THA with ceramic articulation.Materials and MethodsA total of 112 revision THAs were evaluated. The mean age at the time of surgery was 51.6 years (27.7 to 84.2 years). The mean duration of the follow-up periods was 6.3 years (2.3 to 11.4 years).ResultsThe Harris hip scores improved from an average of 56.2 at the index surgery to an average of 93.3 at the last follow-up (P<0.001). None of hips showed osteolysis or ceramic head fracture. One hip showed aseptic loosening in the acetabular component with squeaking that caused a re-revision. There were nine cases of dislocation. The survivorship at 5 years was 94.5% (95% confidence interval, 87.9% to 97.6%) with revision for any reason as the endpoint and 100% with femoral revision.ConclusionThe ceramic articulation is one of good bearing options for revision THA in patients with a long life expectancy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.