Abstract
A CdZnTe (CZT) detector was utilized in mammographic x-ray spectroscopy under clinical conditions. First, the detector response was investigated using γ-rays from 241Am. The escape of secondary (Compton scattered and K fluorescent) x-rays and tailing due to carrier trapping were minor in the mammographic energy range. In addition, the transmission of primary x-rays was minimal from the results calculated using the mass attenuation coefficients of CZT. Therefore, spectral distortion in this energy range was expected to be negligible. Secondly, x-ray spectroscopy was carried out with the CZT detector. The measured spectra were in good agreement with the spectra obtained with the Compton-scatter method with a high-purity germanium detector. Moreover, the half-value layers (HVLs) calculated from the CZT spectra were consistent with the HVLs measured with an ionization chamber. The results indicate that a CZT detector can be utilized in mammographic x-ray spectroscopy without any corrections.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.