Abstract

Design, construction and performance characteristics of the streamer chambers for the central muon detector at CDF are described. A single hit TDC is used for measurements in the drift (azimuth) direction while charge division is used for measurements along the sense wire (pseudorapidity). The chambers operate in the limited streamer mode with a 50%/50% ratio of argon/ethane bubbled through ethanol. Measurements in a cosmic ray test stand, pion test beam and as part of the CDF detector indicate that an rms resolution of 250 μm in the drift direction and an rms resolution of 1.2 mm along the sense wire are attainable.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.