Abstract

BackgroundCD4 T-cell counts are still widely used to assess treatment eligibility and follow-up of HIV-infected patients. The World Health Organization (WHO) prequalification of in vitro diagnostics requested a manufacturer independent laboratory evaluation of the analytical performance at the Institute of Tropical Medicine (ITM) Antwerp, Belgium, of the Muse Auto CD4/CD4% system (Millipore), a new small capillary-flow cytometer dedicated to count absolute CD4-T cells and percentages in venous blood samples from HIV-infected patients.MethodsTwo hundred and fifty (250) patients were recruited from the HIV outpatient clinic at ITM. Accuracy and precision of CD4 T cell counting on fresh EDTA anticoagulated venous blood samples were assessed in the laboratory on a Muse Auto CD4/CD4% system. Extensive precision analyses were performed both on fresh blood and on normal and low stabilized whole blood controls. Accuracy ((bias) was assessed by comparing results from Muse CD4/CD4% to the reference (single-platform FACSCalibur). Clinical misclassification was measured at 500, 350, 200 and 100 cells/μL thresholds.ResultsIntra-assay precision was < 5%, and inter-assay was < 9%. CD4 T cell counts measured on Muse Auto CD4/CD4% System and on the reference instrument resulted in regression slopes of 0.97 for absolute counts and 1.03 for CD4 T cell percentages and a correlation coefficient of 0.99 for both. The average absolute bias as compared to the reference was negligible (4 cells/μL or 0.5%). The absolute average bias on CD4 T cell percentages was < 1%. Clinical misclassification at different CD4 T cell thresholds was small resulting in sensitivities and specificities equal or >90% at all thresholds except at 100 cells/μL (sensitivity = 87%). All samples could be analyzed as there was no repetitive rejection errors recorded.ConclusionsThe Muse Auto CD4/CD4% System performed very well on fresh venous blood samples and met all WHO acceptance criteria for analytical performance of CD4 technologies.

Highlights

  • In 2017, about 36.7 million people were living with HIV of which 20.7 (56.4%) were receiving antiretroviral treatment (ART) [1]

  • Accuracy and precision of CD4 T cell counting on fresh EDTA anticoagulated venous blood samples were assessed in the laboratory on a Muse Auto CD4/CD4% system

  • CD4 T cell counts measured on Muse Auto CD4/CD4% System and on the reference instrument resulted in regression slopes of 0.97 for absolute counts and 1.03 for CD4 T cell percentages and a correlation coefficient of 0.99 for both

Read more

Summary

Introduction

In 2017, about 36.7 million people were living with HIV of which 20.7 (56.4%) were receiving antiretroviral treatment (ART) [1]. The best disease-specific marker to monitor ART in HIV patients is plasma viral load [5]. CD4 remains the best measurement of a patient’s immune and clinical status, risk of opportunistic infections, and it is being used to support diagnostic decision-making, for patients with advanced HIV disease [6]. The World Health Organization (WHO) prequalification of in vitro diagnostics requested a manufacturer independent laboratory evaluation of the analytical performance at the Institute of Tropical Medicine (ITM) Antwerp, Belgium, of the Muse Auto CD4/ CD4% system (Millipore), a new small capillary-flow cytometer dedicated to count absolute CD4-T cells and percentages in venous blood samples from HIV-infected patients

Objectives
Methods
Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call