Abstract

In 1997, the Consultative Committee for Photometry and Radiometry (CCPR) initiated a supplementary comparison of spectral radiance in the wavelength range from 220 nm to 2500 nm (CCPR-S1) using tungsten strip-filament lamps as transfer standards. Five national metrology institutes (NMIs) took part in the comparison: BNM/INM (France), NIST (USA), NRC (Canada), PTB (Germany) and VNIIOFI (Russia), with VNIIOFI as the pilot laboratory. Each NMI provided the transfer lamps that were used to transfer their measurements to the pilot laboratory. The intercomparison sequence began with the participant measurements, then the pilot measurements, followed by a second set of measurements by the participant laboratory. The measurements were carried out from 1998 to 2002, with the final report completed in 2008. This paper presents the descriptions of measurement facilities and uncertainties of the participants, as well as the comparison results that were analysed in accordance with the Guidelines for CCPR Comparisons Report Preparation, and a re-evaluation of the results taking into account the instability of some of the transfer lamps. Excluding a few wavelengths, all participants agree with each other within ±1.5%. The disagreement decreases to approximately ±1.0% when the anomalous data are excluded from the analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.