Abstract

A cation exchange separation of Si from Pu is niade prior to spectrographic determination to reduce the health hazard and spectral interference from Pu. Boiler cap electrodes in conjunction with a sodium carbonate buffer double the sensitivity normally obtained without caps in the same direct current arc methods. The lower limit of detection is 1 ppm Si. The over-all relative standard deviation of the method was within plus or minus 9%. (auth)

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.