Abstract

A decrease of the Tear Meniscus Height (TMH) has been proposed as a useful indicator for Aqueous Deficient Dry Eye (ADDE) categorization. The present study aimed to calculate a TMH cut-off criterion for the categorization or severity assessment of ADDE with the Tearscope. 200 participants with a previous Dry Eye Disease (DED) diagnosis according to TFOS DEWS-II criteria were recruited. TMH by slit-lamp illumination and Lipid Layer Pattern (LLP) with Tearscope were assessed to categorise the participants into the ADDE or the Evaporative Dry Eye (EDE) group. The ADDE group was also subdivided into Mild-moderate ADDE and Moderate-severe ADDE based on TMH with slit-lamp. Additionally, the TMH was measured by Tearscope (TMH-Tc). Receiver Operating Characteristics showed that the TMH-Tc have a diagnostic capability to differentiate between ADDE and EDE participants, and between Mild-moderate or Moderate-severe ADDE, with a cut-off value of 0.159 mm (AUC = 0.843 ± 0.035, p < 0.001; sensitivity: 86.4%; specificity: 75.4%) and 0.105 mm (AUC = 0.953 ± 0.025, p < 0.001; sensitivity: 98.1%; specificity: 80.0%), respectively. The present study proposed a cut-off criterion to differentiate between ADDE and EDE participants, or between ADDE severities through TMH assessed by Tearscope.