Abstract

We report two cases of modular head-neck junction failure involving the Stryker Accolade TMZF stem leading to sudden dissociation of the femoral head from the stem. Both patients presented with mechanical symptoms in the hip followed by pain and hip dysfunction. Disassembly of the head and deformation of the male taper were seen on preoperative radiographs. Intraoperatively, both patients had substantial metallosis with a markedly damaged taper requiring stem revision. We recommend regular clinical and radiographic surveillance of patients with the Stryker Accolade TMZF stem, especially those patients with pain and/or mechanical symptoms.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.