Abstract
The failure of total knee prosthetic components is a rare event, and has been most commonly reported around the polyethylene insert of tibia or the tibial prosthesis. The isolated fracture of a femoral component is a rare event. A review of the literature has reported a very similar failure in the same device in the same region of the medial femoral condyle, implanted in the same year, but in a different geographical location in the UK. Issues with polyethylene fracture and subsequent base plate failure are already documented, but those of the femur have not. We recommend timely revision of total knee arthroplasty where component failure is suspected, as this may prevent further damage.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.