To the Editor. —In the AprilArchives(1983;101:665), Gamel and Eiferman drew attention to corneal burns and cataracts caused by argon laser therapy with the Rodenstock Panfunduscopic lens. When analyzing possible causes, they missed the explanation given by Birngruber et al, 1 which is based on the optics of the Panfunduscope. While a focused beam through a Goldmann lens has its smallest diameter at the retinal plane, a properly focused beam with a Panfunduscope has the narrowest beam distribution just anterior to the cornea. This results in a high energy concentration in the cornea, the iris (if the pupil is not well dilated), and the lens. In the lens, the focal energy concentration is still about 25 times higher than in the retina.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.