Abstract
Purpose: To assess the incidence of cartridge cracks during foldable intraocular lens (IOL) insertion and to determine factors that play a role in the development of these cracks.Setting: Leesburg Regional Ambulatory Surgical Care Center, Leesburg, Florida, USA.Methods: Small incision cataract surgery was performed in 350 consecutive cases. A foldable silicone IOL (Allergan Medical Optics SI-40) was inserted in all cases using the Unfolder- cartridge and 3 viscoelastic agents: sodium hyaluronate (Healon®, Vitrax®) and sodium chondroitin sulfate-sodium hyaluronate (Viscoat®).Results: Cartridge cracks occurred in 52 eyes (14.86%). Almost all cracks (98.1%) occurred in cases in which Healon was used to load the IOL. In each case of a cracked cartridge, there was evidence of the plunger overriding the optic edge.Conclusions: We recommend the use of chondroitin-based viscoelastic agents to load the SI-40 foldable IOL to minimize the risk of cartridge cracks. Modifications in the design of the IOL inserter may eliminate the problem of cartridge cracks.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.