Abstract
The effects of variation of the grating separation in a stretcher on the carrier-envelope (CE) phase of amplified pulses are investigated. By translating one of the telescope mirrors in the stretcher with a piezoelectric transducer, it is found that a 1 mum change of the distance causes a 3.7+/-1.2 rad shift of the CE phase, which is consistent with theoretical estimations. The results indicate that optical mounts used for gratings and telescope mirrors must be interferometrically stable; otherwise their vibration and thermal drift will cause significant phase error. The CE phase drift was corrected by feedback controlling the grating separation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.