Abstract

There are only limited data on the impact of device-related artifacts on image quality in cardiovascular magnetic resonance imaging (CMR) in patients with pacemakers (PM). Adenosine stress perfusion, T1-weighted imaging and flow measurement as well as valve characterization have not been evaluated previously concerning artifact burden. We aimed to assess image quality in all routinely used CMR sequences. We analyzed 2623 myocardial segments in CMR scans of 61 patients with MR conditional PM (mean age 72.1 ± 11.5years), 23 (37.7%) with right sided, 38 (62.3%) with left-sided devices. There were no relevant artifacts in patients with right-sided devices irrespective of the imaging sequence. In left-sided implants no PM-induced artifacts were found in first pass perfusion sequence, flow analysis and T1 weighted imaging. Only few patients with left-sided devices showed significant PM-artifacts in aortic (3/38, 7.9%)/mitral (n = 2/38, 5.3%) valve imaging and STIR (n = 3/35, 8.6%). In STIR only 14/805 (1.7%) segments were involved. In left-sided PM SSFP cine sequences had more artifact burden than LGE with 377/1505 (25.0%) vs. 162/1505 (10.8%) myocardial segments involved by relevant artifacts respectively (p < 0.001). Apart from cine and LGE imaging in anterior myocardial segments with left-sided implants presence of MRI conditional pacemakers does not affect CMR image quality in multimodal CMR examinations to a significant extent. Our data supports evidence that reduced image quality does not need to be a major concern in PM patients undergoing CMR.