Abstract
The purpose of our study was to determine if the rate of lead-related complications was increased with the Medtronic CapSureFix MRI™ SureScan™ 5086 MRI pacing lead (5086; Medtronic Inc., Minneapolis, MN, USA) compared to the previous generation of Medtronic CapSureFix Novus™ 5076 pacing lead (5076). The 5086 lead is a newly introduced active-fixation pacemaker lead designed to be used conditionally in a magnetic resonance (MR) scanner. This lead has specific design changes compared to the previous generation of 5076 pacing leads. This study was a retrospective case control study of 65 consecutive patients implanted with two 5086 leads compared to 92 consecutive control patients implanted with two 5076 leads over a 14-month period at a high-volume tertiary care hospital. Pericarditis, pericardial effusion, cardiac tamponade, or death within 30 days of implant were seen in eight patients from the 5086 cohort and two from the 5076 cohort (odds ratio 6.3, 95% confidence interval 1.3-30.8, P = 0.02). Lead dislodgement occurred in four of the 5086 patients and in none of the 5076 patients (P < 0.03). In a high-volume center, the incidence of pericarditis, cardiac perforation, tamponade, death, and lead dislodgement was significantly higher with the MR-conditional Medtronic 5086 lead when compared to the previous generation Medtronic 5076 lead.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.