Abstract

Optimal contrast enhancement is crucial for the detection of coronary artery stenoses and atherosclerotic changes in coronary CT angiography (CTA). To demonstrate the feasibility of using the cardiac output (CO) obtained from the test bolus injection data-set (COtest) as a factor in contrast injection rate revision of the following coronary CTA. The test bolus injection data-sets of 52 consecutive coronary CTAs were examined. COtest was calculated from the test bolus data-set. Aortic peak enhancement (APE) was measured on the following coronary CTA. We simulated the APE at a fixed contrast injection rate of 4 mL/s (simAPE) in each patient. The ranges of COtest and simAPE were 2.82-7.56 L/min and 194-527 Hounsfield Units, respectively. There was a significant negative correlation (R = -0.802, P < 0.001) between simAPE and COtest. COtest can be used for injection rate revision on coronary CTA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.