Abstract
Equipoise remains about how best to measure cardiac output (CO) in patients with left ventricular assist devices (LVAD). In this study, direct Fick CO was compared with thermodilution (TD) and indirect Fick (iFick) CO in 61 LVAD patients. TD and LaFarge iFick showed moderate correlation with direct Fick (R2=0.49 and R2=0.38, p < 0.001 for both), while Dehmer and Bergstra iFick showed poor correlation with direct Fick (R2=0.29 and R2=0.31, p < 0.001 for both). Absolute bias between all CO estimation techniques and direct Fick CO was lowest for TD compared to iFick methods but significant for all methods. All methods tended to overestimate CO compared to direct Fick, with greatest overestimation present in those with the lowest measured direct Fick CO. Bias and frequency of significant discrepancy were least using TD and Lafarge iFick CO estimation methods in this study, with TD CO demonstrating modestly better correlation and less heteroscedasticity compared to Lafarge.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.